54
SME1202A
8)
Handler Interface
SME1202A Insulation Resistance Meter equips with a Handler interface which is mainly
used to output the sorting result. When the instrument is applied to an automatic
component sorting test system, this interface will output the handshake signal and the
sorting result output signal. The sorting result output corresponds to the comparison result
output of the current comparator bin.
Terminals and Their Descriptions:
Port
No.
Port name
Description
1
/START
Low level pulse, it is valid when in falling edge. When the
instrument is in external trigger mode and this signal is
valid, the instrument will trigger one measurement.
2
/PASS2
BIN2 passes.
3
/FAIL
BIN failed.
4
VCC
Internal electric source, +5V.
5
EXTV
External power supply, +5V.
6
/PASS1
BIN1 passes.
7
/PASS3
BIN3 passes.
8
/EOC
End of the measurement.
9
COM
External power supply ground.
NOTE: the sequence chart below is obtained when the MEAS DISP and BIN DISP
are set as OFF.
Non-programmable Handler sequence chart (OUTP RESU is set as LEVEL)
Programmable Handler sequence chart (OUTP RESU is set as PULSE)