6. Troubleshooting Cont’d.
/
Cycle aborted, NOT STERILE and
CYCLE FAULT messages.
If kinked, straighten the tube. If the tube cannot be
straightened, remove it from the push-in fitting
attached to the Statim. Depress the collar on the
fitting and with the other hand pull firmly on the tube.
Once the tube is free of the fitting, cut the damaged
section of tubing away using a sharp instrument. Be
sure that you leave enough tube to reach the unit when
you re-attach the tube to the exhaust fitting. If the tube
is too short to remove a section, contact your SciCan
dealer for a replacement (Part No. 01-100204S).
Make sure the compressor is working. To check,
remove the exhaust tube from the waste bottle during
the Air Drying phase of a cycle. Wear gloves to do this
as the tubing, fittings, and waste bottle will be hot.
Place the free end into a glass of water. If there is not a
strong, steady flow of bubbles, the compressor is not
functioning properly. Contact your SciCan dealer.
Record the CYCLE FAULT number. Wait a few
minutes and attempt another cycle before proceeding
to the next solution. If the problem persists remove the
cassette. Be careful. The metal parts will be hot and
the cassette will contain hot steam. Inspect the
cassette to ensure that the holes in the back of the seal
are perfectly aligned, and that the flexible lip of the seal
is completely free. Check the exhaust tube for kinks or
obstructions. If kinked, straighten the tube. If the tube
cannot be straightened, remove it from the push-in
fitting attached to the Statim. Depress the collar on the
fitting and, with the other hand pull firmly on the tube.
Once the tube is free of the fitting, cut the damaged
section of tubing away using a sharp instrument. Be
sure that you leave enough tube to reach the unit when
you re-attach the tube to the exhaust fitting. If the tube
is too short to remove a section, contact your SciCan
dealer for a replacement (Part No. 01-100204S).
Check that the Statim has not inadvertently been
exposed to any electrical interference. Refer to the
Installation Section 3.1 Environmental Considerations,
for details.
Try running another cycle. If the problem persists,
record the cycle fault message number and contact
your dealer.