Start-up
24
01.00 | FTR-AXIA | Commissioning instructions | en | 1517835
Command: "set"
The "set" command reports all settings. Many settings are read-
only fields that are configured on the sensor during factory
calibration.
"set" fields
Parameter
Meaning
Other
parameters
serialNum
Display serial number of the product.
partNum
Display number of the calibration part.
calFamily
Display calibration family.
calTime
Display the date of the last calibration.
max0–5
Display maximum nominal value of axes in force/
torque numbers.
forceUnits
Define unit for force measurement.
0 = Lbf
1 = N
2 = Klbf
3 = kN
4 = Kg
torqueunits Define unit for torque measurement.
0 = Lbf-in
1 = Lbf-ft
2 = Nm
3 = Nmm
4 = Kg-cm
5 = kN-m
cpf
Display counts per force.
cpt
Display counts per torque.
peakPos0–
5 /
peakNeg0–5
Display positive / negative peak loads.
sensorHwVe
r
Display version of the sensor hardware.
adcRate
ADC sampling rate in Hertz (488, 976, 1953, 3906 or
7912)
rdtRate
Transmission rate RDT protocol in Hertz (between 1
and adcRate)
rdtSize
Number of RDT transmissions per UDP packet
filTC
Filter selection (0–8)
calib
Select calibration
0
or
1
.
location
Display location of the sensor.
serNum
Display serial number.
hwProdCod
e
Product code of the hardware