SATO 8310-00 Instruction Manual Download Page 8

 

 

Alarm Function 

The alarm is triggered with sound of buzzer when the WBGT reading exceeds the set value. 

Note:

  The alarm function is only available for WBGT. It cannot be set for temperature, globe 

temperature or relative humidity. 

Also, this instrument is automatically turned off if it is not in operation for more than 20 

minutes.  In case of using it for longer than 20 minutes,  release  the Auto-Power Off 

function.   

* Refer to the 

“Auto Power-Off Function” for the instructions. 

 

How to Set Alarm Function 

  Press the MODE/Up key to display the WBGT mode. 

  Press the SET key for more than 2 seconds to display the alarm setting mode. 

  Press the MODE/Up key to increase the value of the blinking digit. 

Press the IN_OUT key to change the digit. 

Note: 

Set the alarm within the range of 20.0 to 49.9°C.   

It is not possible to disable the alarm function. If you do not want the alarm to be 

alerted, set the alarm to its maximum limit, 49.9°C. 

  Press the SET key for longer than 2 seconds to confirm the setting. 

Caution:

  If you turn the power off  before confirming the alarm setting, the instrument 

recalls the last fixed value for the alarm. 

 

Auto Power-Off Function 

If the instrument is not in use for more than 20 minutes, the power is automatically turned off 

for power saving purpose. 

 

How to Release Auto Power-Off Function 

  Make sure the instrument is turned off. 

  Press the POWER key and the HOLD key simultaneously. 

  Release the buttons once the power turns on and the 

“n” mark is shown on the LCD. The 

“n”  mark  disappears  and  the  measuring  mode  starts,  meaning  the  Auto  Power-Off 

function is successfully released. 

Caution   

The  release  setting  is  no  longer  effective  once  the  power  is  turned  off.  Do  the  same 

procedure whenever the Auto Power-Off function is not needed, 

Summary of Contents for 8310-00

Page 1: ...INSTRUCTION MANUAL No 8310 00 Heat Stress Monitor Model SK 150GT...

Page 2: ...this instrument before use for proper usage This instrument does not prevent a heat stress This instrument indicates WBGT index which can be used as guidelines for general heat stress Heat stress is...

Page 3: ...sed battery into a fire Keep battery out of reach of children If swallowed seek a medical attention immediately Dispose the used battery according to your local rules and regulations Please contact us...

Page 4: ...he values manually SET Press this key to set an alarm In WBGT mode press longer than 2 sec to enter the alarm setting mode Press the key longer than 2 sec in alarm setting mode to fix the set value an...

Page 5: ...itor measured data on a PC LCD Display 7 segment Display Display for Temperature TA Globe Temperature TG Relative Humidity RH or WBGT Index WBGT Also it displays the set value in alarm setting mode MI...

Page 6: ...the Low Battery Mark flashing accurate measurement cannot be expected Always use a new battery for replacement How to Measure Expose the sensing section by tuning the cover Turn on the power by pressi...

Page 7: ...ment fully adapt to measuring ambient without condensing before use Handle the globe with great care Damages on the globe or its finishing can have an effect on the measurement accuracy HOLD Function...

Page 8: ...20 0 to 49 9 C It is not possible to disable the alarm function If you do not want the alarm to be alerted set the alarm to its maximum limit 49 9 C Press the SET key for longer than 2 seconds to con...

Page 9: ...n jack on the right side of the unit Connect the 9 pin connector of the cable to the serial port of your computer Turn the power on to active the measurement mode The measurement data will be transmit...

Page 10: ...could be malfunctioning Stop using the instrument and contact the dealer of your purchase or our company for further assistance E33 The internal electrical circuit is malfunctioning Stop using the ins...

Page 11: ...Temperature Thermistor Humidity Electrostatic capacitance type sensor Alarm for WBGT Setting Range 20 0 to 49 9 C in 0 1 C step no deactivation Operation Ambient 0 to 50 C less than 95 rh no condensi...

Page 12: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http www sksato co jp english...

Reviews: