SATO 8310-00 Instruction Manual Download Page 7

 

 

 

Cautions in Measuring WBGT 

The way to calculate WBGT Index differs due to its environment for measurement. 

Make  sure  to  set  the  proper  measuring  ambient  of  the  instrument  according  to  the  actual 

measuring environment. 

  Press the POWER key to turn on. 

  Select the WBGT mode by pressing the MODE/Up key. 

  Press the IN_OUT/NEXT key for longer than 2 seconds to select a measuring ambient. 

  IN 

: When measuring indoor or outdoor without direct sunlight 

  OUT  : When measuring outdoor with direct sunlight 

 

 

CAUTION

 

 

Never use the instrument out of its operating ambient. 

 

Never use the instrument out of its measuring range. 

 

Abrupt  change  of  ambient  temperature  can  effect  on  the  accuracy  of  the  instrument. 

Have the instrument fully adapt to measuring ambient without condensing before use. 

 

Handle the globe with great care. 

Damages on the globe or its finishing can have an effect on the measurement accuracy. 

 

HOLD Function 

When measurement values are unstable, use HOLD function is effective for easy reading. 

  Press the HOLD key in the measurement mode.   

HOLD indicator is lit and the measured value will be held.   

  Press the HOLD key again to release the function. 

HOLD indicator is off and the function will be released. 

 

MAX/MIN Function

 

Maximum and minimum measurement values can be displayed while taking measurement. 

  Press MN/MX/RST key in the measurement mode. The display changes cyclically. 

[MIN]   

[MAX]   

[Measurement Mode] 

 

 

  Press MN/MX/RST key for more than 2 seconds to reset the MIN/MAX values. 

The MIN/MAX values can be cleared by turning the power off. 

The MIN/MAX values are constantly updated in all modes, except for in HOLD mode. 

Summary of Contents for 8310-00

Page 1: ...INSTRUCTION MANUAL No 8310 00 Heat Stress Monitor Model SK 150GT...

Page 2: ...this instrument before use for proper usage This instrument does not prevent a heat stress This instrument indicates WBGT index which can be used as guidelines for general heat stress Heat stress is...

Page 3: ...sed battery into a fire Keep battery out of reach of children If swallowed seek a medical attention immediately Dispose the used battery according to your local rules and regulations Please contact us...

Page 4: ...he values manually SET Press this key to set an alarm In WBGT mode press longer than 2 sec to enter the alarm setting mode Press the key longer than 2 sec in alarm setting mode to fix the set value an...

Page 5: ...itor measured data on a PC LCD Display 7 segment Display Display for Temperature TA Globe Temperature TG Relative Humidity RH or WBGT Index WBGT Also it displays the set value in alarm setting mode MI...

Page 6: ...the Low Battery Mark flashing accurate measurement cannot be expected Always use a new battery for replacement How to Measure Expose the sensing section by tuning the cover Turn on the power by pressi...

Page 7: ...ment fully adapt to measuring ambient without condensing before use Handle the globe with great care Damages on the globe or its finishing can have an effect on the measurement accuracy HOLD Function...

Page 8: ...20 0 to 49 9 C It is not possible to disable the alarm function If you do not want the alarm to be alerted set the alarm to its maximum limit 49 9 C Press the SET key for longer than 2 seconds to con...

Page 9: ...n jack on the right side of the unit Connect the 9 pin connector of the cable to the serial port of your computer Turn the power on to active the measurement mode The measurement data will be transmit...

Page 10: ...could be malfunctioning Stop using the instrument and contact the dealer of your purchase or our company for further assistance E33 The internal electrical circuit is malfunctioning Stop using the ins...

Page 11: ...Temperature Thermistor Humidity Electrostatic capacitance type sensor Alarm for WBGT Setting Range 20 0 to 49 9 C in 0 1 C step no deactivation Operation Ambient 0 to 50 C less than 95 rh no condensi...

Page 12: ...3 4 Kanda kajicho Chiyoda ku Tokyo 101 0045 Japan URL http www sksato co jp english...

Reviews: