12.2 ELECTROMAGNETIC IMMUNITY
I-SURGE + is intended for use in the electromagnetic environment specified below. The I-SURGE + should
only be used in such an environment.
Immunity test
Test level
IEC 60601-1-2
Compliance level
Electromagnetic environment - guidance
Electrostatic
discharge
(ESD) EN61000-4-2
6kV contact
8kV air
6kV
8kV
Floor should be wood, concrete or ceramic
tile. If floor are covered with synthetic
material, the relative humidity should be at
least 30%
Burst/Fast transient
EN 61000-4-4
2kV for power supply
lines
2kV for power
supply lines
Quality of the Main power source should be
that of a typical commercial or hospital
environment
Surge
EN 61000-4-5
1kV differential
mode
2kV common mode
1kV differential
mode
2kV common mode
Quality of the Main power source should be
that of a typical commercial or hospital
environment
Voltage dips, short
interruptions and
voltage variation on
power supply input
lines.
EN 61100-4-11
5%Ut
(
95% dip in U
t
) for 0,5 cycle
40% Ut
(60% dip in U
v
) for 5 cycle
70% U
t
(30% dip in U
t
) per 25 cycle
5% U
t
(
95% dip in U
t
) for 5 sec
5%Ut
(
95% dip in U
t
) for 0,5
cycle
40% Ut
(60% dip in U
v
) for 5 cycle
70% U
t
(30% dip in U
t
) per 25 cycle
5% U
t
(
95% dip in U
t
) for 5 sec
Quality of the Main power source should be
that of a typical commercial or hospital
environment If the user of the I-SURGE +
requires continued operation during power
main interruptions, it is recommended than
the I-SURGE + be powered from an
uninterruptible power supply or battery.
Power frequency
(50/60 Hz)
magnetic field
EN 61100-4-8
3A/m
3A/m
Power frequency magnetic fields should be at
a typical commercial or hospital
establishment level.
I-SURGE + is intended for use in the electromagnetic environment specified below. The I-SURGE + should
only be used in such an environment.
Immunity test
Test level
IEC 60601-1-2
Compliance level Electromagnetic Environment - Guidance
Conducted RF
EN 61000-4-6
Radiated RF
IEC 61000-4-3
3 Veff from 150KHz
to 80MHz
3 Veff from 80MHz
to 2.5GHz
3 Veff from
150KHz to 80MHz
3 Veff from 80MHz
to 2.5GHz
Portable and mobile RF communication
equipment should be used no closer to any part
of the I-SURGE +, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d = 1,2
P from
150KHz to 80MHz
d = 1,2
P from
80MHz to 800MHz
d = 2,3
P from
800MHz to 2.5GHz
where P is the maximum output power rating of
the transmitter in watts (w) according to the
transmitter manufacturer and recommended
separation distance in meters (m).
Field strengths from fixed RF transmitter, as determined by an electromagnetic site survey
a
should be less than the
compliance level in each frequency range
b
. Interference may occur in the proximity of equipment marked with the
following symbol: