d. Technical characteristics
Manufacturer
By Dental – Acteon Group –
Pistoia – Italy
Nominal Power
70 VA
Distributor
SATELEC- ACTEON - France
Fuses
2x1AT(230V) - 2x2AT(115V) 5 x 20
Dimensions
230 x 160 x 200
Protection
Class I
Materials
Aluminum
Type
BF
Micro motor
40000 rpm / Torque: 28.6
mN.m
Class MD
IIa rule 9 in Annex IX 93/42/EEC
Noise
< 65 dBA
Console
IP21
Supply
Voltage
230 VAC/115 VAC
Micro motor
IP65
Frequency
50 Hz /60 Hz
Foot Control
IP21
Not suitable for use in presence of flammable anesthetics or oxygen.
Conditions of use: Temp +18°C/+40°C (+64°F/+104°F) RH < 80%. Altitude restriction:
3.000m (10.000 ft)
Shipping and Storage conditions: Temp +5°C/+65°C (+41°F/+149°F) RH < 20-95% non condensing.
IV– INSTALLATION / FIRST USE
4.1 UNPACKAGING THE DEVICE
Upon reception of the device, check for any damage caused in transit.
Contact your supplier if necessary.
4.2 RECOMMENDATIONS
Have your device connected to the power supply by a certified dental installation technician. The
electrical connection of the I-SURGE + must comply with the standards applicable in your country.
A grounded power supply socket must be used.
4.3 INSTALLATION
Important: Do not place the I-SURGE + close to or on top of another device.
Do not place the power cord and the footswitch cord in a wire cover or in a cable gland.
1.
Remove the unit and its accessories from their original packaging and place on a flat surface. Not
install near heat sources, direct or indirect.
2.
Verify the contents of the box according to the section
“3.1 PHYSICAL DESCRIPTION”.
3.
Do not kink the motor hose since it may damage it.
4.
All accessories are supplied non-sterile with the exception of tubing.
5.
The serial number on the Unit’s rear panel must be the same as the one on transport documents.
6.
Allow at least 150 mm free space around the unit for cooling ventilation.