19
- If necessary, do not use the device and replace any faulty component.
The I-SURGE + is prepared for use as follows:
- Make sure that the device is connected correctly and properly insulated.
-
Micromotor connection:
Connect the micro motor into the receptacle on the front of the central unit (Right –
white – 8 Pins)
(Fig. 2 -10), (Fig. 8).
Connect the “E-type” ISO 3964 contra-angle to the micro motor and fit an instrument
(drill, bur, etc.) to the contra-angle.
Place the micro motor on the motor holder.
-
Footswitch connection:
Connect the footswitch into the receptacle on the front of the central unit (Left – Blue –
5 Pins), (Fig. 2 – 11).
Place the footswitch on the ground.
While inserting the plug, align the guide on the connector in an upright position to fit into the
corresponding female guide on the I-SURGE +. Connectors have a spring retention system to
help avoid accidentally disconnecting the unit's cords and cables.
- Irrigation connection:
- Hang the bottle or bag of physiological saline from the bracket.
- Insert the irrigation line to the pump (“IN” = perforator side ; “OUT” = contra-angle side) and insert
the irrigation line perforating pin into the bottle or bag of physiological saline (Fig. 10).
- Fasten the irrigation line to the cord using line clips.
- Connect the end of the irrigation line to the contra-angle.
Note:
This sequence must be completed in a controlled asepsis context for the safety of your patients.
An operating assistant may be needed.
The manufacturer accepts no responsibility in the case of irrigation tubes other than those
mentioned in this Manual, being used.
- Switch on the device:
- Switch the power switch to position I (power on).
- Make any necessary setting adjustments (irrigation flow rate, speed, torque, program, etc.)
- Use I-SURGE + in accordance with good dental practices.
IX – SHUTTING DOWN THE DEVICE
Upon completion of the dental procedure:
- Remove the irrigation line perforating pin from the bottle or bag of physiological saline.
- Remove the bottle or bag from the bracket.
- Immerse the irrigation line perforating pin in a vessel containing distilled water.
- Rinse the irrigation line and the contra-angle by running the flush function until the recipient and
irrigation lines are completely empty.
- Remove the irrigation line clips and dispose of it in a safety container for soiled medical items.
- Disconnect the irrigation line from the contra-angle and dispose of it in a safety container for soiled
medical items.