26
•PCB damaged.
If you cannot find the cause of the malfunction, immediately discontinue
use of the device and contact dealer, the authorized service center, or the
manufacturer directly.
Fault detected: The Motor will not turn off
Possible cause
Solution
•Keyboard and/or foot pedal damaged Try to turn off the motor by pressing the
[Motor ON/OFF] button
on the
keyboard or by using the foot pedal.
If the motor will not turn off, immediately discontinue use of the device
and contact SATELEC after-sales service.
•Broken connections inside.
•PCB damaged
Immediately discontinue use of the device and contact SATELEC after-sales
service.
•Wrong voltage.
Verify that the voltage matches the power supply. If it does, try to switch
the central unit off and then on again, or connect the device to another
electrical outlet.
Fault detected: The pump is not working property
Possible cause
Solution
•Rotor is blocked.
Disconnect power to the central unit. Check to see if the rotor of the
peristaltic pump rotates correctly and is not blocked by any internal or
external objects. Remove any visible blockage if possible. Otherwise,
contact SATELEC after-sales service.
•Tube not correctly inserted
•Tube damaged.
To verify the proper tube connection follow the instructions on Chapter VIII
– USING THE DEVICE.
Check for damage along the tube, kinks, partial occlusions, etc. Arrange for
immediate replacement of the tube.
Fault detected: The pump stops suddenly
Possible cause
Solution
•Rotor is blocked.
Disconnect power to the central unit. Check to see if the rotor of the
peristaltic pump rotates correctly and is not blocked by any internal or
external objects. Remove any visible blockage if possible. Otherwise,
contact SATELEC after-sales service.
•Pump damaged/blocked.
•PCB damaged.
Immediately discontinue use of the device and contact SATELEC after-sales
service.
XII – ELECTROMAGNETIC COMPATIBILITY
Important:
The power cord, micro-motor cord, and footswitch cord must be kept apart.
The I-SURGE + requires special precautions to be taken with regard to electromagnetic compatibility. It
must be installed and prepared for use as described in chapter IV.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere
with the I-SURGE +. The recommended separation distances in this paragraph must therefore be
complied with.
The I-SURGE + must not be used near to or on top of another device. If this cannot be avoided, its
operation under the conditions of use must be checked beforehand.