c. Torque adjustment
/-
: increases or decreases the torque values (24 steps). Torque values are in Ncm and depend
on contra-angle reduction rate. See the following table with the handpiece 20:1:
Ncm
3.0
6.0
9.0
12.0
15.0
18.0
21.0
24.0
27.0
30.0
33.0
36.0
39.0
42.0
45.0
48.0
51.0
54.0
57.0
60.0
63.0
66.0
69.0
MAX
Torque accuracy is guarantee only with a contrangle correctly maintained. The torque
values are only for handpieces that operate properly. Micro motor stops when set torque is
reached. The torque value in counterclockwise rotation is always = Max Ncm
d. Flow
FLOW
: Selects variable irrigation pump flow rate. See the following table
0 = no irrigation
1 = MIN
2 = MED
3 = MAX
e. Reverse
REVERSE:
Switch between forward and reverse motor rotation (when in reverse, console will emit a
beeping tone).
f. Motor
MOTOR ON/OFF
:
Switches motor ON and OFF (if enabled).
g. Pump
PUMP ON/OFF
:
Switches pump ON and OFF (if enabled).
Always check the values before use. Danger of injury for incorrect use.
h. Footswitch
The most important functions can be activated either by the control panel and the variable-speed
footswitch. (Fig. 4).
VIII– USING THE DEVICE
Before turning on the central unit, check to see that you have completed the installation of
ALL COMPONENTS.
Important:
- Do not disconnect the micro-motor cord when the device is switched on and the footswitch is pressed.
- Do not engage the instrument to or disengage it from the contra-angle when the micromotor is running.
- All accessories must be cleaned, disinfected and sterilized before use.
- For your safety and that of your patient, the I-SURGE + must not be used with accessories other than
those supplied by BY DENTAL and/or SATELEC.
- Check the integrity of the device and its accessories before and after each use in order to detect any
problem.