20
- Remove the instrument fastened to the contra-angle.
- Disconnect the contra-angle from the I-SURGE + micro-motor.
- Disconnect the micro-motor cord from the I-SURGE + unit.
- Switch the device off (O).
X – ROUTINE MAINTENANCE / STERILIZATION
Important: The I-SURGE + micro-motor and cord are not sterile upon delivery. During
sterilization, items made of different metals must not come into contact. Any contact
would establish galvanic couples and cause local damage. To avoid this happening, place
the items in individual sterilizable bags or in a sterilization box.
To maintain the sterility or asepsis of the accessories (contra-angle, micro-motor, micromotor cord,
etc.), store them in sealed bags or containers suitable for medical practice.
The maintenance and/or sterilization instructions below must be applied before each use of the device.
Avoid using cleaning and disinfectant agents containing inflammable substances. If this is not possible,
make sure all such agent has evaporated and that there is no combustible matter on the device or its
accessories before it is switched on.
Sterilized items must be allowed to dry and cool to room temperature before they are reused.
Before sterilization, check that your autoclave is clean and that the water used is of adequate quality.
After each sterilization cycle, remove the items from the autoclave immediately in order to minimize
the risks of metal corrosion.
Cleaning
Materials:
Water, Neutral
detergent, Damp cloth,
Alcohol
To remove dirt or traces of drugs on the surface, wipe with a cloth
slightly dampened with water, neutral detergent or alcohol where
permitted. Be sure to remove any dirt and residue from all
components.
Disinfection
Materials: Disinfectant tested and approved with demonstrable bactericidal, fungicidal and
virucidal properties.
Method: wipe with a cloth slightly dampened with disinfectant products.
Drying
Materials: Soft cloth
Drying carefully with a soft and clean cloth all the components.
Maintenance
The micro motor requires no maintenance. - Check at least every 6 months the cable of the
micro motor. In case of deterioration, it must be replaced at an authorized service center.
Packaging
Before sterilizing the engine, put it in an envelope of Kraft-Polypropylene. The bagging can
maintain the sterility of the material for an extended period of up to 30 days if stored in a dry
and temperature stable environment. Double bagging will increase the time of sterility
maintenance twofold.
Sterilization
Materials: Steam sterilizer according to EN 13060, envelopes of Kraft-Polypropylene.
Parameters: 134°C (273°F) per 5 minutes (2.1-2.2 bars; 30.88 psi).
Shipping and
Storage
conditions
Temp +5°C/+65°C (+41°F/+149°F) RH < 20-95% non condensing.