![Sartorius MA37 Operating Instructions Manual Download Page 58](http://html1.mh-extra.com/html/sartorius/ma37/ma37_operating-instructions-manual_1199627058.webp)
58
Care and Maintenance
Operating Instructions Moisture Analyzer MA37
Contaminated Equipment:
– Health risk from product contamination due to product deposits and the
collection of residue with microbial contamination .
– Health risk from biological or microbiological substances .
– Observe cleaning specifications .
– Check the cleaning result closely .
Cleaning the Heating Module
Danger of Burns
:
The inner part of the heating module and parts of the sample chamber may be
extremely hot .
Wait until the heating module has cooled completely .
Avoid touching the heating element . To remove spray residues or deposits from
the heating element, use a weak solvent such as ethanol .
Unlocking the Heating Module
1) Open the heating module all the way .
2) Pull the unlocking handle at the back of the heating element to unlock the
heating element .
Removing the Heating Module
3) Pull the heating module up out of the guide .