![Sartorius MA37 Operating Instructions Manual Download Page 33](http://html1.mh-extra.com/html/sartorius/ma37/ma37_operating-instructions-manual_1199627033.webp)
33
Moisture Analysis
Operating Instructions Moisture Analyzer MA37
Moisture Analysis
Purpose: The moisture analyzer can be used to determine the moisture content of liquid,
pasty, or solid samples .
Accessories: Depending on the consistency of the sample, one or two filters are recommended in
order to distribute the sample optimally in the sample pan .
– Liquid sample: Tare one filter with the sample pan, then distribute the sample
drop by drop onto the filter .
– Pasty sample: Tare two filters with the sample pan, then place the sample
between the two filters and press together evenly .
– Solid sample: Distribute the granulated sample evenly onto the sample pan
without a filter . Use one or two filters for solid samples that liquefy at high
temperatures or contain fat or sugar, in order to prevent crusting .
Danger of poisoning or caustic burns!
Substances that contain poisonous or caustic ingredients may produce poisonous
gases when dried, which can cause irritation (to the eyes, skin, or airways) or nausea
and lead to serious injury or death .
– These samples must only be dried in a laboratory hood .
Corrosion due to aggressive vapors!
Substances that emit aggressive vapors when heated may cause condensation to
form on cooler housing parts and lead to corrosion .
– Use small sample quantities of these substances .
Viewing and Changing Method Parameters
Moisture in %M
Start new measurement
Standard
Fully automatic
105
°C
Wednesday, 2014-01-08
16:50
Before starting the moisture analysis, you can view the current parameters and
change them if necessary .
– A method configuration for moisture analysis is preset . You can change and
save the parameters to suit your requirements .
The most important parameters (method name, temperature, switch-off criterion)
are displayed .
t
To check or view additional parameters, select the gray button at the bottom of
the main screen on the display .
Customize parameters
No. of filters
Print lines
0
Off
Standby temperature
Off
y
A summary of the current parameters is displayed .
If a lab printer is connected, you can print out the method parameters:
t
Select the print symbol .
y
The parameters are printed out .
t
Select
o
to close the display .
t
To change the parameters, select the menu key at the bottom left of the main
screen on the display .