MA Performance Test
Operating Instructions Moisture Analyzer MA37
MA Performance Test
The performance test can be used to check whether the analyzer has a fault
or is ready for operation . During a performance test, both the heating unit and
the weighing system are tested at the same time . You will need a ReproEasy
pad (accessory) in order to carry out a performance test .
Starting the Performance Test
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
t
Select “MA performance test .”
y
The display switches to a view with three heating temperatures .
Please ensure that the device has not been heated for at least one hour
before a performance test is carried out .
t
MA performance test
The device is too warm for a performance
test.
The device is still cooling down.
m
t
Select “
m
” to confirm the message and start the performance test again later .
MA performance test
Temperature 70 °C
Temperature 100 °C
Temperature 130 °C
t
Select one of the suggested temperatures given for the method you are using .
Open hood
0.000
g
D
t
Open the hood .
y
The next step is displayed when the hood is opened .