44
Calibration and Adjustment
Operating Instructions Moisture Analyzer MA37
Viewing the Calibration Report
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
t
To view the most recent calibration reports: Go to the
Calibration/Adjustment
menu and select the
Calibration report
submenu .
Here you can select the most recent reports for the MA performance tests
or weighing system adjustments and then access the most recent calibration
reports, which are sorted by date .
Up to 99 entries per day are saved . Calibration reports older than 30 days are
deleted .
y
Menu Calibration report .
Example of a Calibration Report:
--------------------
2014-07-29 12:02
Sartorius
Mod. MA37-1
SerNo. 0031710609
BAC: 00-53-01
APC: 01-80-01
DOC: 07-40-03
--------------------
2014-07-29
12:02
External calibration
Start: manually
Set + 50.000 g
Dev - 0.002 g
External adjustment
Dev 0.000 g
Temp + 23.7
°
C
--------------------
Name:
--------------------
Calibration Report
MA performance test
Weighing system adjustment