Practical Example
Weighing in Components
Settings (changes in the factory settings required for this example):
Setup: App: Application 3: Formulation: Automatic storage: On, first value at stability
Setup: App: Application 3: Formulation: Minimum load for automatic storage: 100 digits
Setup: App: Application 3: Formulation: Evaluation mode, MR function: Final evaluation, print
Step
Key (or instruction)
Display/Output
1.
Turn on the balance and configure
e
the settings as indicated above
2.
Delete old formulation data, if necessary
c
3.
Tare the balance
w
4.
Place the empty container
Place load on the balance
on the balance (here: 180.59 g)
5.
Tare the balance
w
6.
Weigh in the first component
Place components
(here: 42.88 g)
in container
7.
Store components in the
M+
soft key
formulation memory
Balance is tared automatically
Components are printed out
automatically
8.
Weigh in the next component
Place components
(here: 50.80 g)
in container
Components are stored in the
totalization memory at stability
and printed out
Balance is tared automatically
9.
Repeat step 7 as required
10. Print final evaluation
MR
soft key
(here: with total weight of all
components: 212.43 g)
11. Delete old formulation data, if necessary
c
s
Comp2 + 50.80 g
s
--------------------
n
2
Tot.cp+
212.43 g
16.01.1997
14:10
--------------------