Error Code/Message Displayed
Cause
Solution/Remarks
Differential weighing:
SAMPLE: Confirm delete/
“SAMPLE: delete/omit” prompt
Select
Yes
to delete
omit
on display page for samples
Select
Omit
to omit
SAMPLE: Include
“SAMPLE: include” prompt
Select
Omit
to include
on display page for samples
a sample already omitted
Cannot store
File manager:
Delete lot(s)
– Not possible to save data
– No available memory
Cannot load
File manager:
Delete lot(s)
– Not possible to load data
– Memory capacity limit reached
Only 30 backweighs
An attempt was made to save a
None
possible
31st backweighing operation
LOT: already exists
Lot already exists on the display
Choose a different lot ID
page for
LOTS
No sample
The
Sample
key was pressed when
Save sample first
the display page for lots was shown,
but there are no samples in the lot selected
Out of range
On the display page for
LOTS
Enter the correct lot or sample ID
or
SAMPLES,
an alphanumeric
lot or samle ID was input and not found
Not enough memory space
An attempt was made to save more
Use less memory or delete one or more lots
or
than 999 samples using the
999 samples maximum
numeric keys and the
#Spl
key
Sample omitted
An attempt was made to save
None
data from an omitted sample
Value too small to accept
An attempt was made to
Place the particular weight on the balance
save a tare, initial or backweight
that is less than one display digit
No choice available
Factor
was selected while attempting
Selection not possible
to activate the display page for results
No 2
nd
resolution available
Contact your local Sartorius Service Center
CF not possible
Only one sample or certain portions
Samples can be deleted one
of a sample can be deleted by pressing
at a time on the display page for samples
the
c
key. This message indicates that
further delete functions are not possible.
Calculated statistics
Message output when statistics are being
Goes out automatically
calculated. This process can take several
seconds if there are many samples.
No statistics available
No valid backweights available in this lot
Goes out automatically
No net initial wts.
In serial and combined weighing,
Measure initial weights
available
no initial weights found
If any other errors occur, contact your local Sartorius Service Center!