88
Practical Example
Totalize counted pieces
Settings (changes in the factory settings required for this example):
Setup: App: Application 1: Counting
Setup: App: Application 3: Totalizing: Evaluated values: Net + calculated
Setup: App: Application 3: Totalizing: Evaluation mode, MR function: Final evaluation, d print
Step
Key (or instruction)
Display/Output
1. Turn on the balance and configure
e
the settings as indicated above
2. Delete old totalization data, if necessary
c
3. Tare the balance
w
4. Toggle to Application 1: Counting
d
5. Place the displayed number of
Place parts to be counted
parts on the balance (here: 10 pcs)
on the balance
6. Initialize the Counting application
Start
soft key
7. Remove the reference sample
Unload the balance
quantity and toggle to Totalizing
d
8. Place a number of parts on the
Place parts on the balance
balance (here: 50 pcs)
nRef 10 pcs
wRef 35.28900 g