Practical Example
Document the amount of evaporation of a sample with defined surface, temperature and air pressure at preset intervals
of 1 minute, 30 seconds.
Settings (changes in the factory settings required for this example):
Setup: App: Application 2: Time-controlled functions
Setup: Balance functions: Taring: Without stability
Setup: Print in weighing mode: Manual/auto print mode: Manual without stability
Step
Key (or instruction)
Display/Output
1. Turn on the balance and configure
e
the settings as indicated above
2. Delete stored values if necessary
c
3. Place container with sample
w
on the balance and tare
4. Enter time interval: 1 minute,
1 . 3 0
30 seconds
5. Store time interval
Interv
soft key
6. Begin documentation
Start
soft key
(Time remaining until the
next printout is displayed
in the text line)
Printout of evaporation amount
every 1
1
minutes
7. Stop the documentation procedure
Stop
soft key