55
Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring (not during alphanumeric input),
– printing,
you can also access the following
functions from this application:
Calibration/Adjustment
G
Press the
Cal
soft key
> See “Calibration/Adjustment” for
further instructions
Toggling to the Next Application
G
Press
d
> See the section on the corresponding
application program for further
instructions
Setup (setting parameters)
G
Press
o
> See “Configuration” for
further instructions
Turning Off the Balance
G
Press
e
> The balance shuts off
Preparation
To calculate a value in percent, the reference percentage must be known. This value can
be entered into the weighing-in-percent program in one of three ways:
– The last reference percentage entered is loaded and displayed when you turn on the
balance. Place the corresponding weight on the balance and initialize the weighing-in-
percent program;
– With automatic initialization switched on (Setup: App: Auto start application when power
goes on: On), the balance goes into the “weighing in percent” mode when you turn it on
and loads the last reference percentage entered as well as the corresponding reference
weight;
– Enter the reference weight using the numeric keys and store it (
Wxx%
soft key).
G
Turn on the balance: Press
e
> Sartorius logo is displayed
G
Select the Weighing in percent application in the Setup menu: Press
o
G
Select
Application parameters
: press the
q
soft key 2
+
,
then the
O
soft key once
G
Select
Application 1 (basic settings)
: press the
O
soft key
G
Select
Percent weigh.
: repeatedly press the
Q
or
q
soft key
G
Confirm
Percent weigh.
: press the
O
soft key
Percent weigh.
Weight storage
ο
Display accuracy
accuracy
+1 decimal place
+2 decimal places
Decimal places
None
for readout in pct.
1 decimal place
ο
2 decimal places
3 decimal places
4 decimal places
5 decimal places
6 decimal places
Display calculated
ο
Residue
value
Loss
Ratio1 (DR)
Ratio2 (OR)
ο
= factory setting
see also the “Application Parameters (Overview)” in the chapter entitled “Configuration”
G
Save settings and exit the Setup menu: press the
oo
soft key
Equations
The following equations are used for the corresponding calculations:
Residue (weighing in percent) = Current weight / 100% weight
+
100%
Loss (percent-DIFF:)
= (Current weight – 100% weight) / 100% weight
+
100%
Ratio1 (percent-Ratio 1:)
= (100% weight – current weight) / current weight
+
100%
Ratio2 (percent Ratio 2:)
= 100% weight / current weight
+
100%