Internal Calibration/Adjustment
First set either
Internal
cal./ adjustment
or
Selection mode
(factory set-
ting) in the Setup menu under “Balance
functions: Calibration/adjustment:
CAL key function.”
Inside the balance housing is a built-in
motorized calibration weight.
The internal calibration/adjustment
sequence is as follows:
G
Select the calibration function:
Press the
Cal
soft key and then the
Start
soft key
> The internal calibration weight is
applied automatically
> The balance is calibrated
> If the setting
Calibrate,
then auto adjust
is selected
in the Balance menu, the balance is now
automatically adjusted
> If the setting
Calibrate,
then manual adjust
is
selected in the Balance menu, you can
end “Internal cab/ adjustment” now;
to start it, press the
Start
soft key
without adjusting the balance (see
“Calibration and Adjustment Sequence”,
next column)
> The internal calibration weight is
removed
> (ISO/GLP/GMP printout: see page 116)
45
Calibration and Adjustment Sequence
In the Setup menu, you can configure
the balance so that:
– calibration is always followed
automatically by adjustment
Calibrate, then auto
adjust
(factory setting) or
– you have the choice of ending
the sequence or starting adjustment
after calibration
Calibrate,
then manual adjust
If no deviation is determined in
calibration, or the deviation is within
the tolerance limits dictated by the
degree of accuracy you require, it is not
necessary to adjust the balance. In this
case, you can end the calibration/
adjustment sequence after calibration.
There are 2 softkeys active at this point:
–
Start
to start adjustment
–
End
to end the sequence