The difference between the weight value and the true mass is displayed, with +/- sign.
External calibration
Calibration record is printed, if adjustment was not performed and the process was
Nom.
+
10000 g
stopped by pressing
(
Diff. +
1 g
)
Activate adjustment (press the
(
key to cancel)
The calibration weight is displayed at the conclusion of calibration
--------------------
A GMP-compliant printout is generated
24.10.2002
10:15
Model
CW3P1-12ED-L
Ser.no.
12345678
Vers.
1.1013.11.2
BVers.
01-26-02
--------------------
External calibration
Nom.
+
10000 g
Diff. +
1 g
External adjustment
Diff. +
0 g
--------------------
24.10.2002
10:15
Name:
--------------------
44