62
§
Connect the cable inside the housing as follows:
– Route the cable through the cable gland.
– Close and tighten the cable gland in accordance with the applicable regulations.
– Strip the casing from a section of the cable end (see illustration). The shielding (1)
must have contact with the clamps (2).
– Expose approx. 15 cm (6 inches) of the individually isolated wires (3) for installation.
– Route the cable through the cable gland.
– Make sure the shield is in contact with the clamps, because the cable is grounded
by the shield.
§
Connecting the wires inside the indicator:
– Expose approximately 5 cm (2 inches) of the isolated wires for installation.
– Remove approximately 1 cm (1/2 inch) of the isolation from the wires and affix ferrules
to the wire ends.
– Connect the wires securely in accordance with the terminal assignments.
§
After you close the housing again, use a pressure gauge to check the integrity of the
IP65/67-protection. For details, contact the Sartorius Service Center.
Cabling Diagram (Adapter Cable for PC)
(CW3P indicator: Adapter cable 7357312; CW3S indicator: Adapter cable YCC02-D9F6).
Diagram for connecting a computer or other peripheral device to the indicator using
the RS-232-C/V24 standard and cables up to 15 m (50 ft.) long:
Cabling Diagrams
Connection assignments for the cable from the indicator to an RS-232 PC interface.
25-pin D-Sub male connector
9-contact D-Sub female connector
(Model CW3P)
1
Sgn GND 7
5 GND
Indicator
TxD
2
2 RxD
side
RxD
3
3 TxD
PC side
DTR
20
8 CTS
CTS
5
4 DTR
Open cable end
9-contact D-Sub female connector
(Model CW3S)
Sgn GND10
5 GND
Indicator
TxD
7
2 RxD
side
RxD
8
3 TxD
PC side
DTR
9
8 CTS
CTS
6
4 DTR
1
Summary of Contents for CH3E
Page 73: ...Dimensions for CH Platforms Scale Drawings CH3G CH3E 73...
Page 80: ...80...
Page 81: ...81...
Page 82: ...82...
Page 83: ...83...
Page 84: ...84...
Page 85: ...85...
Page 86: ...86...
Page 87: ...87...
Page 88: ...88...
Page 89: ...89...
Page 90: ...90...
Page 92: ...92...
Page 94: ......
Page 95: ......