56
Error Codes
Pos. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * E r r * * # # * * * * CRLF
or
S t a t * * * * * E r r * # # # * * * * CRLF
*:
Space
#:
Error code number (2 or 3 digits)
Characters for
ID code I
1
)
Meaning
G
#
Gross value
N
Net value
T
Application tare memory 1
T2
Application tare memory 2
Diff
Difference from calibration value
Nom.
Exact calibration weight
nRef
Reference sample quantity
pRef
Reference percentage
wRef
Reference sample weight
Qnt
Result from Counting (piece count)
and Neutral Measurement
applications
mDef
Target value for Animal weighing
x-Net
Result from Animal Weighing
Setp
Target value for Checkweighing
W.Diff
Absolute difference (e.g., in kg)
in Checkweighing
Lim
Deviation in % in Checkweighing
Max
Upper limit for Checkweighing
Min
Lower limit for Checkweighing
Stat
Status
Class
Classification
Limx
Class limit
D
Percentage (as loss)
Prc
Percentage (as residue)
Wxx%
Reference percentage weight
Cmpxxx
Component xxx
Cont.T
Contents of the tare memory
in Net-total Formulation
Tot. cp
Total weight
in Net-total Formulation
PT2
Preset tare
n
Transaction counter
*G
Sum of gross weights in Totalizing
*N
Sum of net weights in Totalizing
Ser.no
Serial number of the platform
or indicator
1
) depends on scale type; for example, not all units or characters
are available on scales verified for use in legal metrology.
Automatic Data Output (SBI)
You can have results of measurement printed automatically
1
).
You can configure the autoprint function to print at certain
intervals (measured in display updates
2
)) and define whether
printing is dependent on stability of the weighing instrument
3
)
How often the display is updated depends on the operating
status and model of the equipment.
Examples:
N
+
153.00 g
Net weight
Stat
Display blank
Stat
L
Display underload
Stat
H
Display overload
Setting:
1
)
3
) Automatic output without stability
or
Automatic output with stability
Factory setting: Manual after stability;
i.e., automatic data output function off.
2
)
Time-dependent automatic data output:
Intervals:1, 2, 10 or 100 display updates
Factory setting: 1 display update
External Keyboard Functions (PC Keyboard)
Configure under:
Setup: Device parameters: Bar code: External keyboard
The key codes implemented here are specific to the German
keyboard layout. The following alphanumeric characters are
implemented (some require ”Shift“ key):
a - z, A - Z, 0 - 9, <space>, ”,.\+’<>/”$@%/();=:_?*
Function key:
PC keyboard
Combics 3
F1
)
key
F2
(
key
F3
n
key
F4
F5 soft key (far left)
F5
F4 soft key (second from left)
F6
F3 soft key (middle)
F7
F2 soft key (second from right)
F8
F1 soft key (far right)
F9
D
key
F10
D
key, long (> 2 sec) (‘Info’ function)
F11
M
key
F12
k
key
p
key
Return
L
F1 soft key (far right)
Up arrow
F3 soft key (middle)
Left arrow
F4 soft key (second from left)
Down arrow
F2 soft key (second from right)
Right arrow
F1 soft key (far right)
Home
c
key
Backspace
c
soft key
ESC
c
soft key
Summary of Contents for CH3E
Page 73: ...Dimensions for CH Platforms Scale Drawings CH3G CH3E 73...
Page 80: ...80...
Page 81: ...81...
Page 82: ...82...
Page 83: ...83...
Page 84: ...84...
Page 85: ...85...
Page 86: ...86...
Page 87: ...87...
Page 88: ...88...
Page 89: ...89...
Page 90: ...90...
Page 92: ...92...
Page 94: ......
Page 95: ......