41
Error Codes
Error codes are shown on the main display for approx. 2 seconds, after which the program
automatically returns to the weighing mode.
Display
Cause
Solution
No segments appear
No AC power is
Check the AC
on the display
available
power supply
The AC adapter
Plug the AC adapter
is not plugged in
into an electrical outlet
Automatic shutoff
Turn on the scale
is configured
Non-rechargeable/
Replace or recharge
rechargeable
batteries using external
batteries are drained
charger
h
The load exceeds
Unload the weighing pan
the scale capacity
l
Something is touching
Move the object
the weighing pan
that is touching the
weighing pan away
e 01
Data output is not
Change the setting
compatible with
to the compatible one
the input format
in the menu
e 02
Calibration/adjustment
Calibrate/adjust only
condition not met, e.g.:
when zero is displayed
– scale not zeroed
Press [ZERO] to
zero scale display
– scale is loaded
Unload scale
e 08
The scale was zeroed
Only zero the scale
outside the zero range
when in the allowable
zero range of ± 2% of
the maximum capacity
e 09
Taring not possible
Press [ZERO]
when the gross value
to zero the scale
<
zero