1. Place dosimeter stickers on the surfaces on which you want to check
the UV dosage.
2. Leave the room and close the door.
CAUTION:
Never try to start a disinfection cycle with someone in the
room as the motion sensors (PIR) will prevent it.
3. Wait about 10 minutes until the disinfection cycle is complete before
entering the room again (2 min. waiting time for the lights to turn on, and
5 min. disinfection cycle).
4. Enter the room and inspect the dosimeter stickers. To interpret their
colours and corresponding UV dose, use the included reference label.
NOTE:
The dosimeters are sold separately and directly to customers.
It is their responsibility to verify their installation and validate the results
with a Sanuvox representative.
Step by Step Operating Test with Dosimeter
MAINTENANCE
How to Replace Lamp(s)
Replacing lamps on the ASEPT.1X Max is quick and easy. Simply reach the unit with the proper equipement, such as a
stepladder, turn off the power source, open unit cover, remove and dispose of old lamp(s), replace with new Sanuvox branded
UV lamp(s), close unit cover and turn power source back on. See below for detailed step by step instructions.
CAUTION: In order to avoid reduced performance of lamp(s), do not touch lamp glass without gloves.
Step by Step Lamp Replacement
1. Turn OFF the ASEPT.1X Max by pushing the power button on the unit.
STEP 1