20
Instruction Manual for ASEPT.1X Max
1-888-726-8869
www.sanuvox.com
The new ASEPT.1X Max is now ready to operate automatically.
OPERATION
How to Operate your Product
Once installed, activate and test your unit before use. If tests are successful, the ASEPT.1X Max will automatically operate a
disinfection cycle after each visit in the room.
CAUTION:
The suggested operating temperature range is 10 to 50°C (50 to 122°F). Operating the ASEPT.1X Max outside
this range will result in decreased performance.
Step by Step Unit Activation
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
1. Turn ON the ASEPT.1X Max by pushing the power button on the unit
and
confirm that there is proper voltage (120V or 230V)
at the orange
connectors with the black wire.
2. Confirm that the door contact works well: door closed, circuit closed OR
door open, circuit open.
3. Leave the room and close the door.
CAUTION:
Never try to start a disinfection cycle with someone in the
room as the motion sensors (PIR) will prevent it.
4. Wait 2 ½ minutes. The lamps light up after 2 minutes.
5. Open the door. You will notice a blue light shutting down indicating that
the unit was on and automatically turned off when you entered the room,
interrupting the disinfection cycle.
If the tests have been successful, you can now use your new ASEPT.1X
MAX. Please read the safety instructions carefully before first use.