8-2
ECL-210 Operation Manual
8.2. RS-232C Control
The delimiter for RS-232C communication is set in the factory as CR (carriage return). This can be
changed when power to the ECL-210 is initially switched on. For an explanation of this procedure
please see Section 7.1 Power On.
The following table lists the communication conditions that should be set to enable communication
through the RS-232C interface of the ECL-210.
The ECL-210 provides support for RS-232C communication through the SIF connector on the front
panel. The RTDX (transmit) terminal of the ECL-210 should be connected to the receive terminal of
the external device (e.g. computer). The RRDX (receive) terminal of the ECL-210 should be
connected to the transmit terminal of the external device. Please ensure that the handshake line on the
external device is closed. The terminals for the SIF interface (Hirose HR10-7P-6S) are listed below:
1 bit
Parity
None
Flow Control
None
Stop Bit
Baud rate
Data Length
Communication Method
8 bit
Full Duplex
9600 bps
1
N/C
Not used (Do not connect)
2
N/C
Not used (Do not connect)
3
GND
Ground
4
N/C
Not used (Do not connect)
5
RRDX
RS-232C Receive
6
RTDX
RS-232C Transmit
NOTE
This is a SIF connector on the front panel looking from the front.
1
2
3
6
5
4
N/C
N/C
N/C
RTDX
RRDX
GND