7-14
ECL-210 Operation Manual
MSL-8100
MULTI LIGHT SOURCE
CONTROL SYSTEM
STORE
LOCAL
REMOTE
SIF
SHIFT
ON/OFF
mA
nm
mW
ºC
dBm
LD.ON
APC.ON
COH.ON
STATUS
1
LD
APC
COH
CURR
TEMP
POWER WAVE
FINE
CHANNEL
MODE
2
3
4
5
6
7
8
7.8.2. Shutting LD Off
When the LED is not lit in the [LD.ON] line of the Matrix LD to operate, that LD is not
driving. In this condition, the driving circuit shuts off and the protective circuit works. When
the LED is lit, shut the LD off as the following procedure.
1) Select the target unit to shut LD off on MLS. (Refer 9-4. Selecting a unit in MLS)
2) Set the MODE LD using [MODE] key.
3) Press [ON/OFF] key.
4) The current goes down gradually, and the driving circuit works to the end.
5) When the current is 0, the driving circuit shut off and the protective circuit works.
6) LED is turned off.
Shutting LD off takes time because of the internal process. After executing commands, the
status does not change during processing. After all processes of shutting off are finished, the
status changes. It is same as in communication control.
NOTE