7-32
ECL-210 Operation Manual
6) The unit select indicator of MLS turns light on and off.
4) MLS starts the sequence to shut down the unit selected.
5) The power supply of the units shuts down and the unit LED is lit off.
:
:
:
TYPE
OPT
S / N
ECL-210
PMF
001
LF.IN
RF.IN
LD.ON
APC.ON
COH.ON
STATUS
1
LD
APC
COH
CURR
TEMP
POWER WAVE
FINE
MODE
2
3
4
5
6
7
8
CHANNEL
CHANNEL
7) Loosen the unit fixing screw on the left top of unit by rotating, and pull out the unit with
holding the screw.
8) When MLS confirmed the unit pulled put, it comes back to the ordinary condition
It is possible to pull out units with MLS power ON, but be sure to do along with the procedure
of Pulling Unit Out. It may cause system damage.
Be sure to insert new units or attach blank cover for empty slots. If not, it may cause that the
system can generate heat and can be unstable.
NOTE
NOTE