Doc.-ID: 180516-PT10BA-ES-RoughCam IPM1145-(L)_en_rev.01.docx, page 26 of 40
ure 5.12). Avoid skin or clothing contact with the screw threads as they dispose of LOC-
TITE ® 243™ (chemical basis: Dimethacrylatester). It is used to protect the screws from
losing due to shocks, vibrations but also for sealing purposes.
Figure 5.12
– Opening the RoughCam IPxxx
Pull out very carefully the lead flange in a straight manner, ensuring that the board mod-
ule does not tilt. Due to the created lower pressure this might require some additional
effort.
Attention
: The mounting adapter with the heating module, the temperature controller, the
camera module and the optical module are fixed to the cable gland flange. Beware also
of tilting and work very carefully to avoid damaging the components! Avoid skin and
clothing contact with the cylindrical fit, the surface is treated with lubrication paste (oleag-
inous) to protect the surface against frictional corrosion and mechanical strain.
When opening the housing, make sure not to damage or to pollute the Gylon sealing
(blue)! The sealing is not firmly attached to the cable and supply flange and only fixed by
the means of the screw connections.
Attention!
Beware not to damage the surface of bore hole and shaft (fit).
Attention!
Please make sure not to damage housing sealings and to keep them
clean.