1.4 Preventative Maintenance
The S5800E series preventive maintenance consists of inspecting and cleaning the unit, all RF connectors, and
all accessories. To clean the unit’s surface, use a soft, lint-free cloth dampened with water.
If necessary, the unit case can be cleaned with water and a very mild non-solvent based cleaning solution to
remove dirt and light stains.
NOTE: To avoid damaging the display or instrument case, do not use solvents or abrasive cleaners.
Clean the RF connectors and center pins with a cotton swab dampened with denatured alcohol. The fingers of
the N(f) connectors and the pins of the N(m) connectors should be unbroken and uniform in appearance. If you
are unsure whether the connectors are damaged, gauge the connectors to confirm that the dimensions are correct.
The test port cable should be uniform in appearance, and not stretched, kinked, dented, or broken.
1.5 Calibration Requirements & Annual Verification
Local service centers are responsible for instrument maintenance and calibration once a year. To ensure
performance conformity Saluki recommends a periodical maintenance and calibration.
The Cable & Antenna Analyzer is calibrated with OPEN, SHORT, LOAD calibration components. It is
especially important to verify the OSL calibration components periodically to ensure performance conformity.
1.6 ESD Caution
The S5800E series, like other high performance instruments, is susceptible to electrostatic discharge (ESD)
damage. Coaxial cables and antennas often build up a static charge, which may damage the S5800E input
circuitry. S5800E series operators must be aware of the potential for ESD damage and take all necessary
precautions.
It is recommended to discharge the static by connecting a short or load device to the cable or antenna before
connecting them to S5800E. It is important to remember that the operator may also carry a static charge that can
cause damage. Following the practices outlined in the above standards will ensure a safe environment for both