Document Reference: ENG-UMAN-000731
Revision: 5.00
Page 33 of 61
8.12.2
To copy settings from one Program to another
1.
Drag the Program number to the desired Program Slot for the new program.
Note: You should save the programs to a Save Slot before overwriting a program in case
you want to restore the previous one.
8.12.3
To save a Program into the Save Slots
You can create a backup of a Program from the stimulator or save a newly created Program
for later use in the Save Slots. This stores a copy of the Program as a file on the CI. Error!
Reference source not found.).
Note: If a full Save Slot is selected, the existing Program(s) saved in that slot will be
overwritten.
Note: The programs in the Save Slots will only be saved locally to the CI you use to program
the patient. If you use more than one programming tablet in your clinic, the Programs in
the Save Slots will not be available on any other CI.
8.12.4
To restore a Program from the Save Slots
To restore a saved program into the stimulator:
1.
Touch the desired Save Slot and choose Restore.
a.
You may restore all the Programs (up to 4) saved in that slot, a single
Program or a single Stimulation Set (see Section 8.13).
2.
Turn stimulation on to load the Program into the stimulator (individually, for each
of the Programs that you want to restore). See Figure 8.19 for identifying whether a
Program has been saved to the stimulator.
8.12.5
To load a previously saved Program file
To load a saved program file into a Save Slot:
1.
Touch a Save Slot and choose Load From File, locate and select the file then click
Open.
2.
To restore the saved Program from the Save Slot into the stimulator see Section
8.12.4.
8.13
Interleaved Stimulation
A Program may contain several interleaved Stimulation Sets, targeting several areas of
pain. The CPA has the ability to run up to four (4) Stimulation Sets per Program, at the
same time. Refer to Figure 8.20 below for controls related to interleaved stimulation.