Document Reference: ENG-UMAN-000731
Revision: 5.00
Page 20 of 61
4.
Determine the settings for optimum Feedback Control (FC) therapy (see Section
8.10).
5.
Save the Program to the eCLS/CLS memory (see Section 8.12) and proceed to set up
a maximum of four programs.
6.
Use the CPA with the eCLS/CLS to perform neurophysiological measurements of the
patient.
8.2
Connecting the CI to the CST and eCLS/CLS
To program and download data from the eCLS or CLS, communication must first be
established. The CST connects to the CI to communicate using a radio signal transmitted to
the eCLS or CLS.
1.
Connect the CST to the CI with the USB connection (see Figure 8.1).
Figure 8.1: Connecting the CST to the CI
2.
Position the CST near the stimulator and double-click the Evoke Clinical
Programming Application (CPA) icon
to start the program (see Figure 8.2 for
the main screen layout).
Note: The CST can link with the stimulator within a 1.0 m (3.3 ft) radius.
Note: EMI can affect the quality of this wireless link, see Section 4 – Safety Information.