Touch the ‘More…’ field for the next menu.
Show
Neurophysiological
Measurements
Show buttons to execute neurophysiological
measurement programs in place of the stored
programs display.
EPC Mode Change
Allow Enable or disable changing the
enablement/disablement of Feedback
Control (FC) mode from the EPC.
When EPC Mode Change is enabled the FC
mode can be changed from the EPC.
When EPC Mode Change is disabled the FC
mode cannot be changed from the EPC. A
padlock symbol is displayed to indicate that
EPC Mode Change is disabled.
FC enabled
FC disabled.
Soft Reset
Reset stimulator, clear Safe Mode errors,
clear programs and reconnect
Log Setup
Set up therapy log settings:
Size: Normal (1.5 MB) or Research (13 MB)
Duration: 1 day to 1 year.
Erase Space to Store
Data for …
Configure data acquisition for the specified
duration (1, 10 , 60 or 300s).
Start Data Acquisition
Acquire data.
Stop Data Acquisition
Stop acquiring data.
Download Acq Data
Download acquired data from stimulator.
Start Noise Check (6-11,
High) for …
Start noise check between specified
electrodes (e.g.6-11), amplifier gain (e.g.
High) and duration (0.1, 1, 5 s)
Download Noise Check
Data
Download noise check data from stimulator
9.7
Navigation Menu
Tap the Navigation Menu icon for additional CPA functions.
Close Navigation Menu
Quit
Close CPA – stimulation will continue if on
Survey
Go to the Patient Survey screen (Not
available for Clinical Trial use)