Recommended separation distances between portable and mobile RF

communications equipment and the E-CUBE-Plus

The E-CUBE-Plus is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.

The customer or the user of the E-CUBE-Plus can help prevent electromagnetic interference by maintaining a minimum

distance between portable and mobile RF communications equipment (transmitters) and the E-CUBE-Plus as recommended

below, according to the maximum output power of the communications equipment

Rated maximum output

Power of transmitter

(W)

For transmitters rated at a maximum output power not listed above, the recommended separation distance ‘d’ in meters (m) can
be estimated by using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and persons.

150 kHz ~ 80 MHz

d = 1.2 √P

Separation distance according to frequency of transmitter (m)

800 MHz ~ 2.7 GHz

d = 2.3 √P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

80 MHz ~ 800 MHz

d = 1.2 √P

Cables and accessories

Maximum length

Shield

Complies with

Motor cord

Foot controller with cord

AC power cord

1.2 M

1.7 M

1.7 M

Unshielded

Unshielded

Unshielded

RF emissions, CISPR11

Class A / Group 1

Harmonic emissions

IEC 61000-3-2

EN 61000-3-2

Voltage fluctuations / flicker emissions

IEC 61000-3-3

EN 61000-3-3

Electrostatic discharge (ESD)

IEC 61000-4-2

EN 61000-4-2

Electrical fast transient / burst

IEC 61000-4-4

EN 61000-4-4

Surge

IEC 61000-4-5

EN 61000-4-5

Voltage dips, short interruptions

and voltage variations on

power supply input lines

IEC 61000-4-11

EN 61000-4-11

Power frequency (50/60Hz)

magnetic field

IEC 61000-4-8

EN 61000-4-8

Conducted RF

IEC 61000-4-6

EN 61000-4-6

Radiated RF

IEC 61000-4-3

EN 61000-4-3

Description of BASIC SAFETY and ESSENTIAL PERFORMANCE

The E-CUBE-Plus is intended for use in the electromagnetic environment specified above.
The following essential performance of E-CUBE-Plus should be running in such an environment.

1) Frequency of handpiece should be a value between 20 to 30 ㎑.
2) Irrigation volume should be over the 25 ㎖/min during handpiece is operated.

Summary of Contents for E-CUBE PLUS

Page 1: ...INSTRUCTION MANUAL www saeshin com...

Page 2: ...ES PRODUCT COMPONENTS PART NAME OF CONTROL BOX BUTTONS ON OPERATION PANEL LCD WINDOW CONNECTION OF EACH PART OPERATION PROCESS MAINTENANCE OF THE PRODUCT ATTENTION PRODUCT SPECIFICATION ELECTROMAGNETI...

Page 3: ...ainst electric shock Class equipment To avoid the risk of electric shock the product must be connected to a power source with a protective ground Degree of protection against electric shock Type B app...

Page 4: ...ident or failure User has responsibility to operate and maintain the product This product uses a rechargeable battery Ni MH battery Recycling is possible but some countries may not allow processing so...

Page 5: ...aker If patient is taking antibiotics patient needs to consult with doctor before operation Keep away from fire or heat when using spray containing flammable material Manufacturer does not have any re...

Page 6: ...apex measurement starts the motor starts to rotate automatically and when the measurement is finished the motor stops Available only in APEX ENDO mode AUTO SLOW DOWN The rotational speed of the motor...

Page 7: ...S No No Control Box Adaptor and AC Power Cord Angle Handpiece Option Motor Option Motor Stand Foot Controller Battery Including in the control box COMPONENTS Probe Cord Lip Hook File Clip Connecting C...

Page 8: ...or reverse rotation when changing the rotation angle Gear Ratio Button Use for setting the gear ratio Speed Rotation Angle Button Use for setting the speed RPM Press and hold the button can set the sp...

Page 9: ...O mode it shows the load condition of motor during operating 6 Speed Display It shows the preset RPM In reciprocating mode programs 8 and 9 the currently set angle is displayed when selecting the forw...

Page 10: ...8 Adapter connector Foot controller connector Battery cover Probe connector Motor connector CONNECTION OF EACH PART...

Page 11: ...product at consistent temperature location Please insert the pins of motor cord to the holes position of control box carefully 1 Motor Connection 2 Probe Cord Connection Connect the file clip and lip...

Page 12: ...the connecting cord into the file clip holder with 3 o clock or 9 o clock directions and rotate 90 to fix it Connect the connecting cord to the probe cord then insert it into the cord clip Secure the...

Page 13: ...te the angle handpiece and motor pull out the angle handpiece to its axis side 6 File Connection Please insert the file into the chuck of angle handpiece until stop the file To separate the file and a...

Page 14: ...ously applied to the motor it may automatically stop to prevent overheating In this case leave the motor for a while until it cools down When the display returns to normal the product can be used Auto...

Page 15: ...ot controller Motor Button is of motor the motor stop operation by pressing the motor button once more Motor operation can be controlled by pressing and releasing the foot controller 7 After use place...

Page 16: ...ck the length of the measured root canal and the length by taking an x ray OPERATION PROCESS APEX MODE 1 Turn on the control box by pressing the power button over the 3 seconds 2 Press APEX ENDO butto...

Page 17: ...le only in APEX ENDO mode MAINTENANCE OF THE PRODUCT 1 Cleaning Angle Handpiece Rotate the head in a tap water about 10 seconds with maximum speed to remove the remains blood physiologic saline soluti...

Page 18: ...and let it dry for 30 minutes Temperature 132 Time 4 minute Drying time 30 minute Sterilization Moist heat sterilization Maximum sterilization number is 250 times Moist steam sterilization only Otherw...

Page 19: ...right after charging there may be an abnormality Contact the manufacturer The adaptor fuse is blown out Check the connection The adaptor is not connected to the control box Check the connection The a...

Page 20: ...duct 1 Do not put any physical pressure on the adaptor cord motor cord and foot controller cord 6 Product Disposal Follow your country specific law directives standards and guidelines for the disposal...

Page 21: ...ame AC 100 240V 47 63Hz 0 9 0 34A DC 20 V 1 5A Max Sinpro Electronics Co Ltd MPU31 107 1 Control Box 2 Motor 3 Angle 5 Adaptor Name Gear ratio Bur replacement type Bur dimension Chuck type ACL B 51EPB...

Page 22: ...rom that to which the other device s are connected Consult with the dealer Electromagnetic Emissions The E CUBE Plus is intended for use in the electromagnetic environment specified below The customer...

Page 23: ...cle at 8 angles 0 UT for 1 cycle 70 UT for 25 30 cycles 0 UT for 250 300 cycle 8kV contact 15kV air 2kV for power supply lines 1kV for input output lines 1kV differential mode 2kV common mode 0 UT for...

Page 24: ...o AM FM radio broadcasts and TV broadcasts cannot be determined exactly by theory To assess the electromagnetic environment due to fixed radiofrequency emitters an electromagnetic environment measurem...

Page 25: ...objects and persons 150 kHz 80 MHz d 1 2 P Separation distance according to frequency of transmitter m 800 MHz 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 1...

Page 26: ...use or improper use and treatment Use the product with incorrect input voltage AC voltage Drop the device during setting moving using Use the not recommended consumables or accessories Malfunction aft...

Page 27: ...clave Type B This product should not be mixed with other commercial waste for disposal Non sterile Fuse Speed Reciprocating Angle Ratio Button Torque Button Program Button Apical Line Button Volume Bu...

Page 28: ...53 580 0999 1 52 5 Tel 053 587 2345 Fax 053 580 0918 SOM S007S E 2022 01 27 Rev 0 www saeshin com traus saeshin com MERIDIUS MEDICAL EUROPE LIMITED Unit 3D North Point House North point Business Park...

Reviews: