18
1. Please read the notice below before operating the product.
1) Check various connection codes such as motor code, foot controller code, etc., and check that it works
normally when the switch is turned on.
2) Make sure to check the contact part to patient.
3) If not use the adaptor, check the charging volume of battery.
4) Check the sterile condition of autoclavable parts.
2. Please read the notice below during the product use.
1) Check the status of product and patient constantly during the surgery.
2) When product or patient has any problem, stop the operation in condition of patient’s safety or take
proper action.
3) Do not let patient to approach the product.
3. Please read the notice below after operating the product.
1) Do not put any physical pressure on the adaptor cord, motor cord and foot controller cord.
6. Product Disposal
- Follow your country-specific law, directives, standards and guidelines for the disposal of used electrical devices.
ATTENTION
2) Motor
Problem
The handpiece does
not rotate.
Display the “---” and
setting torque alternately
at the torque display part.
(when the use the function
that the handpiece rotates
in forward after rotating
in reverse automatically)
Cause
The motor cord is not connected
to the control box.
Check the connection.
Check the connection.
Contact the manufacturer.
Contact the manufacturer.
Clean the angle handpiece.
Clean or replace the angle handpiece.
The motor cord is broken or
motor is malfunctioning.
The head of angle handpiece is clogged.
The head of angle handpiece is clogged.
The motor cord is broken or
motor is malfunctioning.
Use the foot controller without
the connection.
Remedy