Operating Instructions Ryeco RMM2010 - CNC 

 

 

 

 

 
 
 
 
 
 
3.5 Noise 
If used as intended, a noise level of 70 dB(A) will not be exceeded in the operational environment 

of the measurement device. 

 
3.6 Vibrations 
No health-related vibrations are emitted by the measurement device. 
 
3.8 General safety notices for the WM1 series 
3.8.1 Laser-specific hazards 
The following explanations are provided in accordance with the applicable DIN standards 
DIN EN 60 825-1 (VDE 0837-1):2015-07 "Safety of laser products" and Accident 
Insurance (DGUV) specification 11 "Durchführungsanweisungen Laserstrahlung" ("Operational 
instructions for laser radiation"). 
The positioning laser emits class 2M laser radiation. The available laser radiation is in the 
visible spectral range of 400 nm to 700 nm. For short exposure times (up to 0.25 s), it is harmless 
to the human eye as long as the cross section of the beam is not decreased with optical 
instruments, e.g. magnifying glasses, lenses, telescopes. As long as no optical instruments, that 
decrease the cross section of the beam, are used, the risk of class 2M laser devices is 
comparable to that of class 2 laser devices

If the eye is accidentally exposed to the laser radiation of a class 2 laser device for a short period, 
i.e. exposure time of up to 0.25 s, it will not be at risk. Class 2 laser devices may therefore 
be used without any other safety measures if it can be ensured that the application requires 
neither deliberate eye contact of more than 0.25 s nor repeated eye contact with the laser radiation 
or the laser radiation reflected by the mirror. Generally, the presence of the eyelid closing 
reflex and other aversion responses to protect the eye cannot be assumed. Therefore if class 2 
laser radiation strikes the eye, you should consciously close your eyes and turn your head away 
immediately. 
 
 
 
 
Warning: The laser emits class 2M laser radiation 
Direct contact with the laser beam for a period of more than 0.25 s can damage the eyes. 
• Never look into the direct or reflected laser beam. 

 

• If the laser radiation strikes the eye, consciously close your eyes and turn your head away from 
the beam immediately. 
• No optical instruments, that decrease the cross section of the beam, may be used. 
• The positioning laser must not be removed from its fastening, rotated in another direction or 
modified or manipulated in any other way. 
• The positioning laser must not be used for any other purpose than to position measurement 
objects. 
• The positioning laser should only be switched on when actually required. 
• Any work, that requires the protective housing of the laser system to be removed, must only be 
completed by qualified specialized staff from Ryf Ltd. 
 
 
 
 
 
 
 
 
 
 

 

Summary of Contents for Ryeco RMM2010-CNC

Page 1: ...Operating Instructions Ryeco RMM2010 CNC Operation Instructions CNC Toolmaker Microscope Ryf RMM2010 CNC ...

Page 2: ...nical changes within the scope of the ongoing development of the product described in these operating instructions without prior notice The details figures and descriptions in these operating instructions therefore only serve as indicative information The illustrations provided in these operating instructions also do not necessarily correspond to the scope of delivery or any spare parts supplies T...

Page 3: ...nction related wear of all wear parts installed in the measurement device 2 Intended uses 2 1 General information about the intended use Workshop microscopes from the RMM2010 series are optical or optical coordinate measurement devices and may be used for the following purposes intended use All models Optical coordinate measurement of two dimensional geometries e g points lines distances circles r...

Page 4: ... by personnel who have the necessary training qualifications and authorization and who have been instructed about the specific conditions at the location of the measurement device Clearly specify and regulate personnel responsibilities regarding installation operation maintenance and cleaning Bear responsibility for the operation of the measurement device Ensure that all employees have read and un...

Page 5: ...mmissioning must be carried out by specialized staff at Ryf Operating the measurement device Trained person with relevant specialist knowledge Instructions on operating the measurement device Maintaining the measurement device Trained personnel Must have received training for the relevant task Must have specialist knowledge about measurement devices to complete maintenance Must be aware of the dan...

Page 6: ... e exposure time of up to 0 25 s it will not be at risk Class 2 laser devices may therefore be used without any other safety measures if it can be ensured that the application requires neither deliberate eye contact of more than 0 25 s nor repeated eye contact with the laser radiation or the laser radiation reflected by the mirror Generally the presence of the eyelid closing reflex and other avers...

Page 7: ...k on electric components must only be carried out by an electrician Locally applicable regulations and guidelines for the installation and operation of electrical systems must be observed Before completing any work on live components remove the mains plug to disconnect the power supply and secure against accidental re start Parts of the measurement device on which inspection maintenance and repair...

Page 8: ... on the housing of the measurement device must be avoided Secure the measurement table of the measurement device so that it cannot spontaneously shift or move during transport The measurement table must be fixed secured with the supplied transport securing device during transport also see The measurement device must only be installed on a level sufficiently load bearing and sufficiently wide surfa...

Page 9: ...C CNC model NOTICE Measurement device RMM2010 CNC is designed according to the latest technology and currently applicable rules and regulations The following figure illustrates an example of a RMM2010 CNC with M3 measurement and analysis software POSITION DESIGNATION 1 Housing 2 Measurement table with glass plate 3 Joystick 4 Camera lens and LED incident light illumination 5 Positioning laser diod...

Page 10: ...rol the axes The following figure illustrates an example of the standard scope of supply POSITION DESIGNATION 1 Connection socket 2 Button to switch between creep speed normal speed 3 Axis interlock button 4 Emergency Stop button 5 GoTo position button optional 6 Z direction rotating 2 3 1 4 5 6 ...

Page 11: ...Operating Instructions Ryeco RMM2010 CNC M3 ...

Page 12: ...C 4 6 1 Measurements using transmitted light and incident light All models of the WM1 series can be used for transmitted light and or incident light measurements The following figure provides a schematic overview of both measurement options ...

Page 13: ...brations of the measurement device and heat sources in its surrounding area should also be excluded as these can cause measurement inaccuracies Certain temperature conditions are required to meet the specified length measurement uncertainty 5 3 Ambient conditions for transport and storage Minimum conditions Permissible ambient temperature 5 C 40 C No dust No direct sunlight No rain and no humidity...

Page 14: ...ion may be used for the camera instead of a USB connection Designation X100 1 is retained 6 0 1 Connections on the joystick CNC model The following figure shows an example of the connections on the joystick of a CNC controlled WM1 similar presentation not to scale 6 1 Connections on the measurement computer The following figure shows an example of the connections on a Panel PC similar presentation...

Page 15: ...ecessary and providing the relevant instructions 2 Complete the reference run by moving the measurement table along the X and Y axis using the handwheel and the camera lens along the Z axis using the handhweel The M3 measurement software is available following the reference run 7 1 2 CNC models 1 An input window indicating that a reference run is required appears 2 Check that the Emergency Stop bu...

Page 16: ...d any warranty claims The functional parts inside the device are cleaned as part of the maintenance inspection work completed by the Customer Services team at Ryf AG in Switzerland We recommend that the measurement device is checked maintained and adjusted if required on an annual basis 1 x year Please contact Customer Services to arrange this tel 41 032 654 21 04 The optical components lens conde...

Page 17: ...Operating Instructions Ryeco RMM2010 CNC ...

Page 18: ...up and down several times to distribute the grease evenly If applicable Remove any excess grease that has collected at the ends of the spindle 15 Clean the Y axis spindle 8c with a lint free cloth move up and down the Y axis several times so that any remaining grease residue is easier to remove 16 Slightly grease the Y axis spindle 8c with a brush 17 Move the Y axis fully up and down several times...

Page 19: ... times to distribute the grease evenly If applicable Remove any excess grease that has collected at the ends of the rails 21 Clean both of the Y axis guide rails 8a 8b with a lint free cloth move up and down the Y axis several times so that any remaining grease residue is easier to remove 22 Slightly grease both of the Y axis guide rails 8a 8b with a brush 23 Move the Y axis fully up and down seve...

Page 20: ...Troubleshooting and fault elimination Any Service work must be completed by the Customer Services team at Ryf Ltd as additional alignment or calibration work and special tools are required The following faults can occur on manual and or CNC models ...

Page 21: ...Operating Instructions Ryeco RMM2010 CNC ...

Page 22: ...omponents against mechanical vibrations and damage 9 3 Disposal To avoid unnecessary remnants recycling assemblies and individual parts of Ryeco Ryf Ltd Messtechnik products is preferable to disposal Mechanical or electric replacement parts should be overhauled and reused Operating and auxiliary materials and any replacement parts that can no longer be used should be disposed of safely and in an e...

Page 23: ...Operating Instructions Ryeco RMM2010 CNC ...

Page 24: ...Operating Instructions Ryeco RMM2010 CNC www ryfag ch ...

Reviews: