43
Doc. No.
:
P30001PE0101-07
Starting at 50 cycles before the 5000-cycle limit, the instrument
will display this screen every time the instrument is switched on.
“XXXX” indicates the number of cycles remaining.
If this number is “0”, the UV lamp must be replaced.
Contact Precision System Science or a distributor for UV lamp replacement.
9.3.
Setup Screen
Press the “4” key from the MENU Screen to display the Setup Screen.
Press the
“ESC” key from the Setup Screen to return to the MENU Screen.
9.3.1.
Setting the date
Press
“1” in the System Setup menu. The following screen appears:
Set the day, month, and year. Press
“SHIFT” and the down arrow to move
the cursor to the right, from DD (day) to MM (month) to YYYY (year).
Press
“SHIFT” and the up arrow to move the cursor to the left, from YYYY to MM
to DD.
Press the up arrow or down arrow to increase or decrease the value
of the selected field.
After setting the date, store the setting by pressing “ENT”.
Press “ESC” to leave the date unchanged.
9.3.2.
Setting the time
Press
“2” in the System Setup menu. The following screen appears:
DD MM YYYY HH:MM
START:Run 1:UV
2:Man 3:Test 4:Setup
Key:START,1,2,3,4
SYSTEM SETUP
1:Date 2:Time
3:Serial Port4:PM
Key:1,2,3,4,ESC
SETUP: DATE
DD MM YYYY
Key:
Up,Dn,SHIFT,ENT,ESC