39
Doc. No.
:
P30001PE0101-07
9.1.2.
Return Tip Screen
Press the
“2” key from the Manual Screen to display the following screen:
Press
“START” to perform the Return Tip operation to return any tips that are
attached to the Syringe unit into tip holders. Press
“ESC” to return to the previous
screen.
9.1.3.
Cleaning Screen
Use the Clean operation to access the piercing unit for cleaning.
The door must initially be closed. Press ”3” in the manual operation screen
to select
the “clean” operation. The following screen appears:
Press the
“START” key to clean the Piercing Unit. After the key is pressed,
the Piercing Unit will be lowered for cleaning, and the following screen appears:
Open the door, and clean the piercing unit, using a paper or cloth moistened with
water or 70% ethanol, depending on how dirty the piercing unit is.
After cleaning the piercing unit, close the door and press the
“ENT” key.
•
Wear a mask, goggles, and gloves for your safety when cleaning the piercing unit.
MANUAL OPERATION
home axis Axis
Executing...
MANUAL OPERATION
return tip
START:Run
Key:START,ESC
MANUAL OPERATION
Clean piercing unit
START:Run
Key:START,ESC
MANUAL OPERATION
Open door and
clean piercing unit
ENT:Done
Key:START,ESC
Caution