9.2.
UV Screen
At the end of a protocol run, a message will be displayed providing the option to
start a UV decontamination run.
You can also manually start a UV decontamination run by switching on
the UV lamp as described below.
Note
UV decontamination helps to reduce possible pathogen contamination
of the magLEAD 12gC stage surfaces.
The efficiency of inactivation must be determined for each specific
organism and depends on the layer thickness and sample type.
Precision System Science cannot guarantee complete eradication of
specific pathogens.
9.2.1.
Switching on the UV lamp
Make sure that the instrument
door is closed. In the MENU Screen, press “1” to
select the UV light function.
The following screen appears:
Use the keys “0” through “9” to set the duration of the decontamination time.
The minimum time is 30 minutes, and the maximum time is 330 minutes.
The default is 30 minutes.
After setting a valid time, press the
“ENT” key. The following screen appears:
If you enter an invalid time (<30 min. or >330 min.),
the following screen will appear:
DD MM YYYY HH:MM
START:Run 1:UV
2:Man 3:Test 4:Setup
Key:START,1,2,3,4
Decontamination
Set Time: 30 min.
Key:0-9,ENT,ESC
Decontamination
Time: 30 min.
Run:START
Key:START,ESC
UV DECONTAMINATION
Number must be
between 30 and 330
Key:ESC