Scan Overview
Background
The Scan function is used to automatically bin groups of up to
100 components. The associated pins in the handler interface are
active when the Scan function is activated.
There are a total of 6 outputs, 3 inputs as well as a GND and
power (+5V) pin.
Interface and pin
assignment
25-Pin D-SUB
(Female)
HANDLER / SCAN / EXT I/O
Pin assignment Relay
Controls the relay output.
Pass
Pass signal. Indicates the compare result is
IN(pass).
Low
Low signal. Indicates a LO compare result.
High
High signal. Indicates a HI compare result.
Clock
The clock signal will pulse high when each group
of output signals (Relay, Pass, Low, High) are
ready. There are up to 100 groups of output
signals
STRB
After all (100) output groups are ready, the STRB
signal will pulse high.