RM-804 & RM-805 Instruction Manual/English
72
13/10/2016 Version No. 001
1. Select Trigger
Delay setting
From one of the main screens, press the
ESC
key so that the menu system at
the bottom of the display has focus.
Go to Meas. Setup and press Enter.
Go to Trigger Delay and press Enter.
Meas. Setup
menu icon
Move
Select menu
or setting
Enter
2. Trigger Delay
setting appears
Use the arrow keys to turn Trigger Delay on and set the delay
time. Press Enter to confirm the settings.
Trigger Delay
setting
Trigger Delay
OFF, ON: 0 ~ 1000ms
Note
Pressing ESC before pressing ENTER will exit the Trigger Delay
settings.
Trigger Edge
Background
The Trigger Edge setting sets the external trigger edge as rising
or falling. By default the trigger edge is set to rising.