MEASUREMENT
13/10/2016 Version No. 001
51
Binning Function
Background
The Binning function is used to grade DUTs into eight different
bins according to 8 sets of upper and lower limits. Two compare
modes can be used in this function, ABS and
△
% modes.
Binning function
indicator
Upper and lower
limits for the 8
bins
Grading results
Reference, compare mode,
beep mode and display mode
1. Select the
Binning function
Press the
Binning
key to access this function.
2. Select the
compare mode
Use the arrow keys to go to the Mode setting.
Press Enter to toggle between ABS or
△
% compare modes.