COMMAND OVERVIEW
13/10/2016 Version No. 001
171
Syntax
Query Syntax
TEMPerature:AMBient:STATe <NR1> | {OFF|ON}
TEMPerature:AMBient:STATe?
Parameter/ Return
parameter
<NR1>
0:OFF.
1:ON.
OFF
Disables the user-set ambient temperature.
ON
Enables the user-set ambient temperature.
Example
TEMP:AMB:STAT OFF
Disables the user-set ambient temperature.
TEMPerature:COMPensate:COEFficient
Set
Query
Description
Sets or returns the temperature coefficient for temperature
compensation function.
Syntax
Query Syntax
TEMPerature:COMPensate:COEFficient <NR1>
TEMPerature:COMPensate:COEFficient?
Parameter/ Return
parameter
< NR1>
-9999~+9999
Example
TEMP:COMP:COEF 3930
Sets the temperature coefficient to 3930ppm.