MEASUREMENT
13/10/2016 Version No. 001
75
1. Select
Ambient
Temperature
setting
From one of the main screens, press the
ESC
key so that the menu system at
the bottom of the display has focus.
Go to Meas. Setup and press Enter.
Go to Ambient Temperature and press
Enter.
Meas. Setup
menu icon
Move
Select menu
or setting
Enter
2. Ambient
Temperature
setting appears
Use the arrow keys to set the Ambient Temperature. Press Enter
to confirm the setting.
Ambient
Temperature
Ambient Temperature
Off, On: -50ºC ~ 399.9ºC
Note
Pressing ESC before pressing ENTER will exit the Ambient
Temperature settings.
Line Frequency
Background
The Line Frequency setting selects the appropriate line filter to
reduce the influence of the AC line frequency on the milliohm
measurements. This setting is set to AUTO by default