RM-804 & RM-805 Instruction Manual/English
166
13/10/2016 Version No. 001
Syntax
SYSTem:LOCal
Parameter
<None>
SYSTem:MDELay:DATa
Set
Query
Description
Sets or returns the measurement delay time.
Syntax
Query Syntax
SYSTem:MDELay:DATa <NRf>
SYSTem:MDELay:DATa?
Parameter/ Return
parameter
<NRf>
0.000~100.000
Unit:ms
For values under 1s, the unit resolution is
1ms.
For values above 1s, the unit resolution is
0.1s.
Example
SYST:MDEL:DAT 1.105
Sets the delay time of measure is 1.1s.
SYST:MDEL:DAT?
>001.100
Returns the measurement delay as 1.1s.