4.3.7
Default database reload
1
Disconnect the DAU power supply.
2
Remove the program EEPROM from the socket, see Figure 4-31.
Figure 4-31. Database loader
3
Insert the DB-loader EEPROM into the EEPROM socket.
4
Power up and wait for about 30 seconds. When the database loading is
finished, all segments on the local display light up:
Comment:
flashing display segments indicate that the data-
base loading failed.
5
Disconnect the power supply.
6
Remove the DB-loader and reinsert the program EEPROM.
7
Power up.
8
Configure the DAU.