Rosemount Model 8732C Integral Mount Magnetic Flowmeter System
4-8
DIGITAL OUTPUT
Pulse Output Scaling equates one transistor switch closure pulse to a
selectable number of volume units. The volume unit used for scaling
pulse output is taken from the numerator of the configured flow units.
For example, if gal/min had been chosen when selecting the flow rate
unit, the volume unit displayed would be gallons.
NOTE
The scaled pulse output is designed to operate between 0 and 1000 Hz.
The electronics will not accept a conversion factor that would result in a
pulse frequency outside that range. The minimum conversion factor
value is found by dividing the upper range value (in units of volume per
second) by 1000 Hz.
When selecting pulse output scaling, remember that the maximum
pulse rate is 1000 Hz. With the 110 percent overrange capability, the
absolute limit is 1100 Hz. For example, if you want the Model 8732C to
pulse every time 0.01 gallons pass through the flowtube, and the flow
rate is 1000 gal/min, you will exceed the 1000 Hz full-scale limit:
The best choice for this parameter depends upon the required
resolution, the number of digits in the totalizer, the extent of range
required, and the maximum counter input frequency.
NOTE
When totalizing on the LOI, the maximum total value is 1,000,000,000.
Auxiliary Output Control
In lieu of the pulse output on the digital terminals, an Auxiliary
Output Control is available to indicate a reverse flow or zero flow
condition. The two terminals are actually a transistor switch closure,
which must be externally powered.
Reverse Flow Enable
This choice activates the switch closure with a reverse flow. A forward
flow is defined by the flow direction arrow on the flowtube. This also
enables the totalizer to count in the reverse direction.
No Flow Enable
This choice activates the switch closure whenever the flow rate drops
below the low flow cutoff.
1000 gal/min
60 sec/min
0.01 gal/pulse
×
---------------------------------------------------------------------------
1666.65 Hz
=
HART Fast Keys
1, 4, 3, 2
LOI
Detailed Setup
Summary of Contents for 8732C
Page 1: ...Model 8732C Integral Mount Magnetic Flowmeter System 00809 0100 4725 English Rev BB ...
Page 2: ......
Page 4: ......
Page 10: ...vi ...
Page 39: ...2 27 Installation ...
Page 40: ...Rosemount Model 8732C Integral Mount Magnetic Flowmeter System 2 28 ...
Page 66: ...Rosemount Model 8732C Integral Mount Magnetic Flowmeter System 4 20 ...
Page 74: ...Rosemount Model 8732C Integral Mount Magnetic Flowmeter System 5 8 ...
Page 80: ...Rosemount Model 8732C Integral Mount Magnetic Flowmeter System 6 6 ...
Page 96: ...Rosemount Model 8732C Integral Mount Magnetic Flowmeter System 8 6 ...
Page 108: ...Rosemount Model 8732C Integral Mount Magnetic Flowmeter System A 12 ...
Page 116: ...Rosemount Model 8732C Integral Mount Magnetic Flowmeter System B 8 ...