Section
6-1
6
Options
OVERVIEW
This section contains descriptions of the options available with the
Models 3144 and 3244MV Smart Temperature Transmitters. The
available options enhance operation and facilitate various installation
configurations.
SAFETY MESSAGES
Instructions and procedures in this section may require special
precautions to ensure the safety of the personnel performing the
operations. Information that potentially raises safety issues is indicated
by a warning symbol ( ). Please refer to the following safety messages
before performing an operation preceded by this symbol.
Warnings
Custom Transmitter
Configuration
(Option Code C1)
Option Code C1 allows you to specify the following data in addition to
the standard configuration parameters.
Trim to Specific
Rosemount RTD
Calibration Schedule
(Transmitter-Sensor
Matching)
(Option Code C2)
Option Code C2 allows you to order the transmitter trimmed to a
specific calibration schedule. This option requires that you specify a
Rosemount Series 65, 68, or 78 RTD sensor ordered with a special
calibration schedule. An additional special curve will be added to the
sensor input choices. For information on ordering sensors calibrated to
specific calibration schedules, refer to the Rosemount Sensors Product
Data Sheet, publication no. 00813-0100-2654 (Volume 1) or publication
no. 00813-0101-2654 (Volume 2).
Five-Point Calibration
(Option Code C4)
Option Code C4 specifies that the transmitter be verified at the the
following five-point calibrations: 0, 25, 50, 75, and 100% analog and
digital output points.
Explosions could result in death or serious injury:
•
Do not remove the transmitter cover in explosive atmospheres when the circuit
is alive.
•
Before connecting a HART communicator in an explosive atmosphere, make
sure the instruments in the loop are installed in accordance with intrinsically
safe or non-incendive field wiring practices.
•
Both transmitter covers must be fully engaged to meet explosion-proof
requirements.
Option Code C1 Data
Date
day, month, year
Descriptor
16 alphanumeric characters
Message
32 alphanumeric characters